Flowmeter, General Information

mig-e8875989e6a19de2.gif

Flowmeters are durable acrylic that will give long-lasting and trouble-free operation, with just a few precautions. The flowmeter is used to measure airflow. To ensure proper accuracy, mount the unit in a vertical position. Use a 9/16 inch wrench on the hex when tightening the piping to the 1/8 inch FNTP threads. Flowmeters are calibrated to +/- 5% accuracy.

On start-up, open the needle valve slowly to purge any entrapped air or gas. The flow is read at the center of the spherical float. Begin flow at a rate equal to the cubic feet of your cabinet. This will purge the cabinet once per hour.

If float visibility becomes hampered, you may need to occasionally clean the unit. Since our flow units have rod guided floats for stability, an occasional cleaning may be necessary if the ball float gets stuck on the rod. Disassemble the flowmeter carefully. Clean the tapered tube with a mild liquid detergent and a soft brush, along with the other parts removed. Rinse all parts with clear water only and dry thoroughly. Avoid all strong solvents, such as benzene or acetone. Reassemble by reversing the disassembly procedure. A small amount of grease (halocarbon or equivalent) on the O-rings will help the seal and prolong the life of the O-ring.

Caution: These flowmeters are designed for use with nitrogen and CDA, at pressures up to 100 PSI and ambient temperature. Do not exceed these limits. Use beyond these limits may cause failure that could result in personal injury.