Installation Instructions For NS16R Nitro-Save® Humidity Control Retro-Fit Kit

mig-a5924a44dcab3a9f.webp

Parts Included

  • Monitor/Solenoid
  • Wire Set
  • AC Adapter
  • Screws for Unit, 4 pcs.
  • Wire Ties
  • Air Tubing
  • Elbow
  • Acrylic Plug
  • 6-32 Plug Tap
  • #36 Drill Bit
  • 3/16″ to 1/2″ Step Drill Bit
  • Manual and Calibration Certificate

Do Not Exceed 100 PSI

Instructions

  1. Remove flowmeter elbow.
  2. Determine the location for your Nitro-Save® unit (as close to center as possible). For even number compartments, place the Nitro-Save® on the second compartment down from the top (as shown on diagram). For odd numbered compartments, place the Nitro-Save® on the middle compartment.
  3. Using the step drill bit, drill a 7/16″ hole for the Nitro-Save® sensor probe.
  4. Drill and tap bracket holes for the monitor (#36 drill bit and 6-32 tap).
  5. Mark the location for your solenoid on the side of the plenum chamber as shown.
  6. Remove the cabinet from the stand and place it on a flat surface.
  7. Remove the back wall.
  8. Drill a 1/2″ hole in the sidewall for the solenoid (step drill bit 1/2″).
  9. Attach the solenoid.
  10. Clean the debris from the plenum chamber and re-install the back wall.
  11. Hold the monitor against the sidewall with the probe inserted into the hole.
  12. Attach with screws.
  13. Connect the wires as shown in the diagram.
  14. Trim the air tubing to the correct size and attach the tubing and elbow.
  15. Use wire ties as needed to neatly hold wires and tubing.
mig-357cd14628ab8ed9.jpg
mig-3d92c9844d34bd83.jpg
Shop Nitro-Save