Installation Instructions For NS16R Nitro-Save® Humidity Control Retro-Fit Kit
Parts Included
- Monitor/Solenoid
- Wire Set
- AC Adapter
- Screws for Unit, 4 pcs.
- Wire Ties
- Air Tubing
- Elbow
- Acrylic Plug
- 6-32 Plug Tap
- #36 Drill Bit
- 3/16″ to 1/2″ Step Drill Bit
- Manual and Calibration Certificate
Do Not Exceed 100 PSI
Instructions
- Remove flowmeter elbow.
- Determine the location for your Nitro-Save® unit (as close to center as possible). For even number compartments, place the Nitro-Save® on the second compartment down from the top (as shown on diagram). For odd numbered compartments, place the Nitro-Save® on the middle compartment.
- Using the step drill bit, drill a 7/16″ hole for the Nitro-Save® sensor probe.
- Drill and tap bracket holes for the monitor (#36 drill bit and 6-32 tap).
- Mark the location for your solenoid on the side of the plenum chamber as shown.
- Remove the cabinet from the stand and place it on a flat surface.
- Remove the back wall.
- Drill a 1/2″ hole in the sidewall for the solenoid (step drill bit 1/2″).
- Attach the solenoid.
- Clean the debris from the plenum chamber and re-install the back wall.
- Hold the monitor against the sidewall with the probe inserted into the hole.
- Attach with screws.
- Connect the wires as shown in the diagram.
- Trim the air tubing to the correct size and attach the tubing and elbow.
- Use wire ties as needed to neatly hold wires and tubing.